Trace oxygen gas analyzer

Model: SY-HG3300-DH

Typical application:

> Measurement of trace oxygen in production process of nitrogen and carbon dioxide

> Measurement of trace oxygen in production process of hydrogen

> Measurement of trace oxygen in Protective Gases for Semiconductor Manufacturing

> Measurement of trace oxygen in inert gas when welding metal materials such as alloy steel and titanium

> Measurement of trace oxygen in Nitrogen and Inert gas in Air Separation unit

> Measurement of trace oxygen in carbon dioxide gas in air separation unit

> Measurement of trace oxygen in Ammonia gas during Metallurgical Industry Quenching Process

> Measurement of trace oxygen in carbon dioxide in the food industry

Enquiry

Principle

Ion flow oxygen technology:

Oxygen molecules in the measured environment diffuse through the diffusion holes to the negative electrode and flow from the negative electrode through the electrolyte oxygen pump to the positive electrode under the action of the electric field, and then return from the positive electrode to the measured environment. The current flowing through the loop has a corresponding ion current value under different oxygen concentration, there is a function relationship between the ion current value and the oxygen concentration in the measured gas. High resolution, good stability, small drift, low power consumption.

Technical data

 

Measured range

0 ~ 10 ppm,  0 ~ 100 ppm,  0 ~ 1000 ppm  O2

Drift

≤1%FS /7d (Long run can be less than 0.5%FS)

Repeatability

0.5%FS

Linearity

1% FS

Power supply

AC220V ± 10%, 50 / 60Hz

Power consumption

≤ 60W

Display screen

5" 640*480 pixel color TFT touch screen

Analog output

0/4 – 20mA  ( RL ≤ 500 Ohm )

Communication

RS485 ( optional )

Alarm output

2 normally open relays ( passive output),

Contact rate: 5A/30VDC;  10A/250VAC

  

 Ambient condition

Ambient temperature

5 ~ 40℃

Ambient humidity

≤ 90% RH

Airflow velocity

0.5m/s

Placing request

horizontal placement

Mechanical vibration

No strong vibration

Electromagnetic interference

No strong Electromagnetic interference

  

Gas-sample condition

Gas-sample temperature

5 ~ 40℃

Gas-sample pressure

2 ~ 50 Kpa

Gas-sample flow rate

30 ~ 60 L/h

Gas-sample moisture content

Before entering the instrument, the gas-sample should be dried. After drying, the dew point temperature should be less than 5°C .

Gas-sample dust content

Before entering the instrument, the gas-sample must be filtered out of dust.

 

Instrument size and weight

Instrument size

488 * 177 * 432 mm

Dimension for mounting

436 * 180 mm

Instrument weight

18 kg