{"id":3161,"date":"2022-05-30T15:17:49","date_gmt":"2022-05-30T07:17:49","guid":{"rendered":"https:\/\/www.hginstrument.com\/?p=3161"},"modified":"2022-05-30T15:17:49","modified_gmt":"2022-05-30T07:17:49","slug":"common-faults-and-treatment-of-liquid-turbine-flow-meter","status":"publish","type":"post","link":"https:\/\/www.hginstrument.com\/it\/common-faults-and-treatment-of-liquid-turbine-flow-meter\/","title":{"rendered":"Guasti comuni e trattamento del misuratore di portata a turbina liquida"},"content":{"rendered":"<table width=\"1123\">\n<tbody>\n<tr style=\"background-color: #d8d8d8;\">\n<td style=\"text-align: center;\" width=\"290\">Fenomeno di guasto<\/td>\n<td style=\"text-align: center;\" width=\"381\">Motivo<\/td>\n<td style=\"text-align: center;\" width=\"452\">Misure di trattamento<\/td>\n<\/tr>\n<tr>\n<td style=\"vertical-align: middle;\" width=\"290\">There is no display when the fluid flows normally.<\/td>\n<td width=\"381\">1. The power line, fuse, etc. is broken or in bad contact.<br \/>\n2. Poor internal contact in display instrument.<br \/>\n3. The coil is broken<br \/>\n4. The sensor&#8217;s flow channel is faulty.<\/td>\n<td width=\"452\">1. Check with the ohmmeter.<br \/>\n2. Adopt the &#8220;back up board&#8221; method to check.<br \/>\n3. Check whether coil is broken or solder spot is fall off.<br \/>\n4. Remove foreign body from the sensor and clean or replace damaged parts.<\/td>\n<\/tr>\n<tr>\n<td style=\"vertical-align: middle;\" width=\"290\">Flow rate display drops gradually<\/td>\n<td width=\"381\">1. Filter blockage<br \/>\n2. The valve spool on the sensor&#8217;s pipe section becomes loosen and the valve opening decreases.<br \/>\n3. Sensor impeller is blocked by debris or foreign bodies enter the bearing clearance, resistance increases.<\/td>\n<td width=\"452\">1. Clean the filter.<br \/>\n2. Judge whether the valve spool is loosen through the valve handwheel adjustment is effective<br \/>\n3. Remove the sensor to eliminate the debris and recalibrate if necessary.<\/td>\n<\/tr>\n<tr>\n<td style=\"vertical-align: middle;\" width=\"290\">The fluid is not flowing, but the flow rate display is not zero<\/td>\n<td width=\"381\">1. Transmission line&#8217;s shielding grounding is not good.<br \/>\n2. The impeller shakes accompany with the pipe vibration.<br \/>\n3. Stop valve is not closed tightly.<br \/>\n4. Display unit&#8217;s internal circuit board or electronic components damage.<\/td>\n<td width=\"452\">1. Check whether shielding wire is good grounding.<br \/>\n2. Reinforce the pipeline, or install brackets at inlet and outlet of the sensor to prevent vibration.<br \/>\n3. Repair or replace the stop valve.<br \/>\n4. Take the &#8220;short-circuit method&#8221; or check item by item to justify the interference source and find out the fault point.<\/td>\n<\/tr>\n<tr>\n<td style=\"vertical-align: middle;\" width=\"290\">The difference between the display value and the empirical value is significant<\/td>\n<td width=\"381\">1. The flow channel of the sensor is faulty.<br \/>\n2. Back pressure of the sensor is insufficient, occur cavitation, which affects impeller rotation.<br \/>\n3. Reason of flowing in pipeline.<br \/>\n4. Internal failure of display unit.<br \/>\n5. The permanent magnet elements in the detector loses the magnetization.<br \/>\n6. The actual flow which pass the sensor exceeds the specified range.<\/td>\n<td width=\"452\">1-4. Find out the cause of the failure and take countermeasures.<br \/>\n5. Replace the elements which loses the magnetization.<br \/>\n6. Select the appropriate sensor.<\/td>\n<\/tr>\n<\/tbody>\n<\/table>","protected":false},"excerpt":{"rendered":"<p>Guasti comuni e trattamento del misuratore di portata a turbina liquida<\/p>","protected":false},"author":1,"featured_media":2768,"comment_status":"closed","ping_status":"closed","sticky":false,"template":"","format":"standard","meta":[],"categories":[2130],"tags":[],"acf":[],"_links":{"self":[{"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/posts\/3161"}],"collection":[{"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/comments?post=3161"}],"version-history":[{"count":0,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/posts\/3161\/revisions"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/media\/2768"}],"wp:attachment":[{"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/media?parent=3161"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/categories?post=3161"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.hginstrument.com\/it\/wp-json\/wp\/v2\/tags?post=3161"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}