Pengukur Aliran Elektromagnetik Sanitasi HGD
Product name: Pengukur Aliran Elektromagnetik Sanitasi HGD
The operation temperature of HGD Sanitary Electromagnetic Flow Meter usually not more than 120℃, fluid’s conductivity not less than 5μS/cm.
HGD Sanitary Electromagnetic Flow Meter is applicable for the industry of high health level requirement, such as pharmaceutical, food industry etc.
Tag Populer: Tentang Suyi Group,Instrumen Aliran,Pemancar Tingkat,pengukur aliran air
Deskripsi Produk
Technical data of HGD Sanitary Electromagnetic Flow Meter
● Applications: Drinking water, fruit juice, milk, Wine, etc.
● Meter size: DN15 – DN150
● Process connection: Tri-clamp
● Tube material: SS304, SS316
● Electrode: SS316L, Hastelloy B, Hastelloy C, Titanium, Tantalum, Pt/Iridium Alloy, SS with Painting Tungsten Carbide
● Liner: PTFE, FEP, Neoprene, Polyurethane, Hard Rubber, PFA
● Measure accuracy : 1.0%, 0.5%, 0.3%;
● Digital display: instantaneous flow and accumulative flow
● Power supply: 24VDC / 220VAC / 3.6V battery
● Output: 4-20mA, pulse, frequency (RS485, HART, Profibus-DP optional)
Wiring Terminal instruction of HGD Sanitary Electromagnetic Flow Meter’s compact converter as follow:
| Terminal kabel | Deskripsi | Keterangan |
| L 2 | Hubungkan 220VAC / 24VDC - | |
| L 1 | Hubungkan 220VAC / 24VDC + | |
| COM | Tanah Keluaran Saat Ini untuk Pengukuran Aliran | Gunakan untuk pengkabelan Output Arus analog |
| I + | Output Arus untuk Pengukuran Aliran | |
| COM | Frekuensi / Output Pulsa Ground | Gunakan untuk pemasangan kabel frekuensi / output pulsa |
| P+ | Output Frekuensi / Pulsa untuk Aliran Dua Arah | |
| AH | Output Alarm untuk Batas Atas | Gunakan untuk pemasangan kabel Output alarm |
| AL | Output Alarm untuk Batas Bawah | |
| COM | Ground Output Alarm (sambungkan dengan Ground Output Arus atau Ground Output Frekuensi/Pulsa) | |
| FUSE | Sekering untuk Catu Daya | |
| T + | Terminal komunikasi (RS485 A+) | Gunakan untuk pemasangan kabel komunikasi |
| G | Tanah Komunikasi RS232 | |
| T - | Terminal komunikasi (RS485 B-) |
Wiring Terminal instruction of HGD Sanitary Electromagnetic Flow Meter’s remote converter as follow:
| Terminal kabel | Deskripsi | Keterangan |
| SIG1 | Sinyal l | Untuk transduser tipe jarak jauh |
| SGND | Tanah Sinyal | |
| SIG2 | Sinyal 2 | |
| DS1 | Terlindung Mengasyikkan 1 | |
| DS2 | Terlindung Mengasyikkan 2 | |
| EXT + | Arus "+" yang Menggairahkan | |
| EXT - | Arus yang Menggairahkan "-" | |
| VDIN | Titik sambungan 24VDC (untuk output arus 2 kabel) | Gunakan untuk pengkabelan Output Arus analog |
| IOUT | Output Arus Analog | |
| ICOM | Tanah Output Arus Analog Ground | |
| POUT | Output Frekuensi Aliran (Pulsa) | Gunakan untuk pemasangan kabel frekuensi (pulsa) Output |
| PCOM | Frekuensi (Pulsa) Keluaran Ground | |
| ALMH | Output Alarm Batas Atas | Gunakan untuk pemasangan kabel Output alarm |
| ALML | Output Alarm Batas Bawah | |
| ACOM | Tanah Keluaran Alarm | |
| TRX+ | Terminal komunikasi (RS485 A+) | Gunakan untuk pemasangan kabel komunikasi |
| TRX- | Terminal komunikasi (RS485 B-) | |
| TCOM | Tanah Komunikasi RS232 | |
| L | Hubungkan 220VAC / 24VDC + | |
| N | Hubungkan 220VAC / 24VDC - |
Parameters list of HGD Sanitary Electromagnetic Flow Meter
| Item | Parameter words | Function description | Setting way | Jangkauan | Password Grade |
| 1 | Language | Display language | Select | Chinese / English | 2 |
| 2 | Comm Addres | Instrument Communication Address | Setting Value | 0 ~ 99 | 2 |
| 3 | Baud Rate | Instrument Communication Speed | Select | 300 ~ 38400 | 2 |
| 4 | Snsr Size | Pipe line Size | Select | 3 ~ 3000 | 2 |
| 5 | Flow Unit | Measuring Unit | Select | L/h, L/m, L/s, m3/h m3/min,m3/s | 2 |
| 6 | Flow Range | Measure range Setting | Setting Value | 0 ~ 99999 | 2 |
| 7 | Flow Rspns | Measure Damping Time | Select | 1 ~ 50 | 2 |
| 8 | Flow Direct | Measuring Direction Selection | Select | Forward / Reverse | 2 |
| 9 | Flow Zero | Flow Zero Amendment | Setting Value | 0 ~ ±9999 | 2 |
| 10 | Flow Cutoff | Small Signal Resection Point | Setting Value | 0 ~ 599.99% | 2 |
| 11 | Cutoff Ena | Resection Point Indication
Permission |
Select | Permit /Prohibit | 2 |
| 12 | Total Unit | Flow Totalize Unit | Select | 0.001m3 ~ 1 m3
0.001 L ~ 1 L |
2 |
| 13 | SegmaN Ena | Reverse Output Permission | Select | Permit /Prohibit | 2 |
| 14 | Analog Type | Current Output Type | Select | 0 ~ 10 mA / 4 ~ 20mA | 2 |
| 15 | Pulse Type | Pulse Output Mode | Select | Frequency/Pulse | 2 |
| 16 | Pulse Fact | Pulse
Unit equivalent |
Select | 0.001m3 ~ 1 m3
0.001 L ~ 1 L |
2 |
| 17 | Freque Max | Frequency Output Range | Select | 1 ~ 5999 Hz | 2 |
| 18 | Mtsnsr Ena | Empty pipe Alarm Permission | Select | Permit/Prohibit | 2 |
| 19 | Mtsnsr Trip | Empty pipe Alarm Threshold | Setting Value | 59999% | 2 |
| 20 | Alm Hi Ena | Upper Limit Alarm Permission | Select | Permit/Prohibit | 2 |
| 21 | Alm Hi Val | Upper Limit Alarm Value | Setting Value | 000.0 ~ 599.99% | 2 |
| 22 | Alm Lo Ena | Lower Limit Alarm Permission | Select | Permit/Prohibit | 2 |
| 23 | Alm Lo Val | Lower Limit Alarm Value | Setting Value | 000.0 ~ 599.99% | 2 |
| 24 | Sys Alm Ena | Excitation Alarm Permission | Select | Permit/Prohibit | 2 |
| 25 | Clr Sum Key | Total Flow Zero Password | Setting Value | 0 ~ 99999 | 3 |
| 26 | Snsr Code1 | Sensor Code 1 | User Setting | Factory Year Month (0 ~ 99999) | 4 |
| 27 | Snsr Code2 | Sensor Code 2 | User Setting | Product Serial No ( 0 ~ 99999) | 4 |
| 28 | Field Type | Exciting frequency type | Select | Type 1, 2, 3 | 4 |
| 29 | Sensor Fact | Sensor Factor Value | Setting Value | 0.0000 ~ 5.9999 | 4 |
| 30 | Total Point | Batch control volume set | Setting Value | 00000~59999 | 4 |
| 31 | Line CRC Ena | Flow amend permission | Select | Permit/Prohibit | 5 |
| 32 | Lineary CRC1 | Flow amend point 1 | User Setting | setting according to the velocity | 5 |
| 33 | Lineary Fact 1 | Flow amend value 1 | User Setting | 0.0000~1.9999 | 5 |
| 34 | Lineary CRC2 | Flow amend point 2 | User Setting | setting according to the velocity | 5 |
| 35 | Lineary Fact 2 | Flow amend value 2 | User Setting | 0.0000~1.9999 | 5 |
| 36 | Lineary CRC3 | Flow amend point 3 | User Setting | setting according to the velocity | 5 |
| 37 | Lineary Fact 3 | Flow amend value 3 | User Setting | 0.0000~1.9999 | 5 |
| 38 | Lineary CRC4 | Flow amend point 4 | User Setting | setting according to the velocity | 5 |
| 39 | Lineary Fact4 | Flow amend value 4 | User Setting | 0.0000~1.9999 | 5 |
| 40 | FwdTotal Lo | Forward Total Flow Low Level | Revisable | 00000 ~ 99999 | 5 |
| 41 | FwdTotal Hi | Forward Total Flow High Level | Revisable | 0000 ~ 9999 | 5 |
| 42 | RevTotal Lo | Reverse Total Flow Low Level | Revisable | 00000 ~ 99999 | 5 |
| 43 | RevTotal Hi | Reverse Total Flow High Level | Revisable | 0000 ~ 9999 | 5 |
| 44 | PlsntLmtEna | Peak Restriction Permission | Select | Permit/Prohibit | 5 |
| 45 | PlsntLmtVal | Peak Restriction Factor | Select | 0.010 ~ 0.800 m/s | 5 |
| 46 | Plsnt Delay | Peak Restriction Time | Select | 400 ~ 2500 ms | 5 |
| 47 | Pass Word 1 | Password grade 1 | Use Revisable | 00000 ~ 99999 | 5 |
| 48 | Pass Word 2 | Password grade 2 | Use Revisable | 00000 ~ 99999 | 5 |
| 49 | Pass Word 3 | Password grade 3 | Use Revisable | 00000 ~ 99999 | 5 |
| 50 | Pass Word 4 | Password grade 4 | Use Revisable | 00000 ~ 99999 | 5 |
| 51 | Analog Zero | Current Zero Amendment | Setting Value | 0.0000 ~ 1.9999 | 5 |
| 52 | Anlg Range | Current Full Span Amendment | Setting Value | 0.0000 ~ 3.9999 | 5 |
| 53 | Meter Fact | Calibration Factor | Setting Value | 0.0000 ~ 5.9999 | 5 |
| 54 | MeterCode 1 | Instrument Code 1 | Factory Setting | Factory Year Month (0 ~ 99999) | 6 |
| 55 | MeterCode 2 | Instrument Code 2 | Factory Setting | Product Serial No. (0~99999) | 6 |
Jalur Perakitan Bengkel
Garis Kalibrasi Bengkel
Gudang
Produk terkait
PERTANYAAN YANG SERING DIAJUKAN
Answer : Silakan kirimkan pertanyaan atau pertanyaan Anda ke email kami: sales@hginstrument.com, kami akan meneruskan pertanyaan tersebut ke teknisi terkait.
Answer : Tolong berkomunikasi dengan bagian penjualan tentang prosedur pengembalian sebelum mengembalikan produk.
Answer : Waktu pengiriman tergantung pada jumlah pesanan dan spesifikasi produk. Jika spesifikasi produk adalah konvensional (tidak istimewa), waktu pengiriman adalah 5 atau 7 hari kerja untuk jumlah pesanan ≤ 10 buah; waktu pengiriman adalah 2 atau 3 minggu kerja untuk jumlah pesanan 15 ~ 30 potongan.
Answer melalui Kurir, Laut atau Udara
Answer : Ya, Umumnya garansi satu tahun.
Answer : Dari tahun 2015 hingga tahun 2021, penjualan tahunan rata-rata adalah RMB 350 juta (sekitar USD 54,7 juta)