Medidor de flujo ultrasónico Doppler montado en pared ( Sensor de flujo tipo inserción )

Principle: use Doppler frequency shift to measure the flow velocity, it is suitable for measuring the fluid which contains solid particles or bubbles, not suitable for measuring pure water. The application scope includes: natural river course, artificial river course, farmland channel, city sewer, enterprise sewage outlet, sewage treatment plant and so on.

Descripción del producto

Features:
● Suitable for pipe sizes ranging from 65 to 4000 mm
● Insertion type sensor, pipe must be fully filled with liquid
● The fluid must include a certain amount of air bubbles or suspended solid particles
● Excellent low flow rate measurement ability, low to 0.05 m/s
● A wide range of flow measurement, high flow rate can reach 12 m/s
● High-temperature transducer is suitable for liquids with temperature from -35℃ ~ 150℃
● The insertion type flow transducer can be installed when the process is at running status.
● User-friendly configuration
● 4-20mA, Relay and OCT outputs
● Accuracy: 2.0% F.S.

 

Applications:
● Raw sewage
● Activated sludge
● Ground water
● Pulp and paper slurry
● Chemical slurry
● Drainage
● Mining re-circulation

 

Technical data of Transmitter:

Measurement principle Doppler ultrasonic
Resolution 0.25 mm/s
Repeatability 0.2% of reading
Precisión 0.5% — 2.0% F.S.
Response time 2-60s for optional
Flow Velocity Range 0.05- 12 m/s
Liquid Types Supported Liquids containing 100 ppm of reflectors and at least 20% of the reflectors are larger than 100 micron.
Power Supply 85-265V AC,   24V DC/500 mA
Enclosure type Montaje en pared
Degree of protection IP66
Operating temperature  -20℃ to +60℃
Material de la carcasa Fiberglass
Measurement Channels 1
Mostrar 2 line × 8 characters LCD, 8-digit rate or 8-digit total (resettable)
Units User Configured (English and Metric)
Rate Rate and Velocity Display
Totalizer gallons, ft³, barrels, lbs, liters, m³,kg
Comunicación 4-20mA, Relay and OCT output
keypad 4  buttons
Talla 244(h)*196(w)*114(d)mm
Weight 2.4 kg

 

Technical data of  Transducer:

Transducers Type Insertion
Degree of protection IP67 or IP68
Suited Liquid Temperature Standard. Temp.: -35℃~85℃
High Temp.: -35℃~150℃
Pipe diameter range 65-4000 mm
Transducer Size 58(h)*58(w)*199(d)mm
Material of transducer Stainless steel
Cable Length Standard configure:  2 * 5m

 

 

PREGUNTAS FRECUENTES

Answer : Envíe su pregunta o consulta a nuestro correo electrónico: sales@hginstrument.com, remitiremos las preguntas al ingeniero correspondiente.

Answer : Por favor comuníquese con el departamento de ventas sobre el procedimiento de devolución antes de devolver el producto.

Answer : El plazo de entrega depende de la cantidad del pedido y de la especificación del producto. Si la especificación del producto es convencional ( no especial ),  El plazo de entrega es de 5 ó 7 días laborables para cantidad del pedido ≤ 10 piezas;  El plazo de entrega es de 2 ó 3 semanas laborables para cantidad del pedido 15 ~ 30 piezas. 

Answer por correo, por mar o por aire

Answer : Sí, generalmente la garantía es de un año.

Answer : De 2015 a 2021, el promedio de ventas anuales es de 350 millones de RMB (unos 54,7 millones de USD).